Sensor |
Range (Input and Output) |
Resolution |
Accuracy / 1 year (Measurement) |
Accuracy / 1 year (Simulation) |
Pt50 (α = 3851) |
-220°C to +1200°C |
0.01°C |
0.012% RDG + 0.06°C |
0.014% RDG + 0.18°C |
Pt100 (α = 3851) |
-220°C to +1200°C |
0.01°C |
0.012% RDG + 0.05°C |
0.014% RDG + 0.12°C |
Pt100 (α = 3916) |
-200°C to +510°C |
0.01°C |
0.012% RDG + 0.05°C |
0.014% RDG + 0.12°C |
Pt100 (α = 3926) |
-210°C to +850°C |
0.01°C |
0.012% RDG + 0.05°C |
0.014% RDG + 0.12°C |
Pt200 (α = 3851) |
-220°C to +600°C |
0.01°C |
0.012% RDG + 0.12°C |
0.014% RDG + 0.33°C |
Pt500 (α = 3851) |
-220°C to +1200°C |
0.01°C |
0.012% RDG + 0.07°C |
0.014% RDG + 0.18°C |
Pt1000 (α = 3851) |
-220°C to +1200°C |
0.01°C |
0.012% RDG + 0.05°C |
0.014% RDG + 0.08°C |
Ni100 (α = 618) |
-60°C to 180°C |
0.01°C |
0.012% RDG + 0.03°C |
0.014% RDG + 0.08°C |
Ni120 (α = 672) |
-40°C to +205°C |
0.01°C |
0.012% RDG + 0.03°C |
0.014% RDG + 0.08°C |
Ni1000 (α = 618) |
-60°C to +180°C |
0.01°C |
0.012% RDG + 0.03°C |
0.014% RDG + 0.08°C |
Cu10 (α = 427) |
-50°C to 150°C |
0.01°C |
0.012% RDG + 0.18°C |
0.014% RDG + 0.1°C |
Cu50 (= 428) |
-50°C to +150°C |
0.01°C |
0.012% RDG + 0.06°C |
0.014% RDG + 0.15°C |
Resistive probes measurements in 2, 3 or 4 wires: automatic recognition of number of connected wires, with indication on screen
Accuracies are given for 4-wire mounted probes
Take into account particular error of temperature sensor used and implementation conditions
Measuring current: 0.25 mA (Measurement) or from 0.1 to 1 mA (Emission)
Temperature coefficient: < 10% of accuracy /°C
Thermocouples: Measurement and simulation
Type |
Input range |
Resolution |
Accuracy / 1 year (Measurement) |
Output range |
Resolution |
Accuracy / 1 year (Simulation) |
K |
-250 to -200°C -200 to -120°C -120 to +0°C 0°C to +1372°C |
0.2°C 0.1 0.05°C 0.05°C |
0.8°C 0.25°C 0.1°C 0.013% RDG + 0.08°C |
-240 to -50°C -50 to -0°C +0°C to +1372°C |
0.2°C 0.1 0.05°C |
0.6°C 0.1°C 0.013% RDG + 0.08°C |
T |
-250 to -200°C -200 to -120°C -120 to -50°C -50 to +400°C |
0.2°C 0.05°C 0.05°C 0.05°C |
0.7°C 0.25°C 0.1°C 0.013% RDG + 0.08°C |
-240 to -100°C -100 to -0°C +0 to +400°C |
0.2°C 0.05°C 0.05°C |
0.4°C 0.1°C 0.013% RDG + 0.08°C |
J |
-210 to -120°C -120 to +0°C +0 to +1200°C |
0.05°C 0.05°C 0.05°C |
0.25°C 0.09°C 0.013% RDG + 0.07°C |
-210 to -0°C +0 to +1200°C |
0.05°C 0.05°C |
0.2°C 0.013% RDG + 0.07°C |
E |
-250 to -200°C -200 to -100°C -100 to -0°C +0 to +1000°C |
0.1°C 0.05°C 0.05°C 0.05°C |
0.45°C 0.15°C 0.07°C 0.013% RDG + 0.05°C |
-240 to -100°C -100 to +40°C +40 to 1000°C |
0.1°C 0.1°C 0.05°C |
0.25°C 0.1°C 0.013% RDG + 0.05°C |
R |
-50 to +150°C +150 to +550°C +550 to 1768°C |
0.5°C 0.2°C 0.1°C |
0.8°C 0.013% RDG + 0.35°C 0.013% RDG + 0.2°C |
-50 to +350°C +350 to +900°C +900 to 1768°C |
0.5°C 0.2°C 0.1°C |
0.5°C 0.013% RDG + 0.35°C 0.013% RDG + 0.2°C |
S |
-50 to +150°C +150 to +550°C +550 to +1768°C |
0.5°C 0.2°C 0.1°C |
0.80°C 0.013% RDG + 0.35°C 0.013% RDG + 0.25°C |
-50 to +120°C +120 to +450°C +450 to +1768°C |
0.5°C 0.2°C 0.1°C |
0.8°C 0.013% RDG + 0.35°C 0.013% RDG + 0.25°C |
B |
+400 to +900°C +900 to +1820°C |
0.2°C 0.1°C |
0.013% RDG + 0.4°C 0.013% RDG + 0.2°C |
+400 to +850°C +850 to +1820°C |
0.2°C 0.1°C |
0.013% RDG + 0.4°C 0.013% RDG + 0.2°C |
U |
-200 to +660°C |
0.05°C |
0.15°C |
-200 to +600°C |
0.05°C |
0.15°C |
L |
-200 to +900°C |
0.05°C |
0.2°C |
-200 to +900°C |
0.05°C |
0.2°C |
C |
-20 to +900°C +900 to 2310°C |
0.1°C 0.1°C |
0.25°C 0.013% RDG + 0.15°C |
-20 to +900°C +900 to 2310°C |
0.1°C 0.1°C |
0.25°C 0.013% RDG + 0.15°C |
N |
-240 to -190°C -190 to -110°C -110 to +0°C +0 to +1300°C |
0.2°C 0.1°C 0.05°C 0.05°C |
0.5°C 0.15°C 0.08°C 0.013% RDG + 0.06°C |
-240 to -190°C -190 to -110°C -110 to +0°C +0 to +1300°C |
0.2°C 0.1°C 0.05°C 0.05°C |
0.3°C 0.15°C 0.08°C 0.015% RDG + 0.06°C |
Platinum |
-100 to +1400°C |
0.05°C |
0.3°C |
-100 to +1400°C |
0.05°C |
0.3°C |
Mo |
+0 to +1375°C |
0.05°C |
0.013% RDG + 0.06°C |
+0 to +1375°C |
0.05°C |
0.013% RDG + 0.06°C |
NiMo/NiCo |
-50 to +1410°C |
0.05°C |
0.013% RDG + 0.30°C |
-50 to +1410°C |
0.05°C |
0.013% RDG + 0.3°C |
Thermocouples: D and G, for specifications, refer to the instruction manual (Available on request)
Accuracy is given for reference @ 0°C.
When using the internal reference junction (except couple B) add an additional uncertainty of 0.2 °C at 0 °C.
It is possible (thermocouple B excepted) to choose by programming the cold junction localization: External at 0°C, internal (temperature compensation of instrument’s terminals) or manually entered.
Temperature coefficient: <10% of accuracy /°C
Display unit: °C and F
Specifications and performances in pressure @23°C ±5°C
Pressure: Measurement by external digital sensor
Range |
0-1 bar |
0-3 bar |
0-10 bar |
0-30 bar |
0-100 bar |
0-300 bar |
0-1000 bar |
Absolute |
X |
X |
X |
X |
X |
X |
X |
Relative |
X |
X |
X |
X |
|
|
|
Available in relative, absolute and differential pressure.
Connector: ¼ gas
Resolution: 0.02% FS
Accuracy:
-0.05% FS from 10 to 40°C
- 0.1% FS from -10 to +10°C and from 40 to 80°C
This digital pressure module ACL433 is connected to CALYS 75 through RS485 serial cable to the digital input connector. All data are digital. Measurements are compensated in temperature by a polynomial correction implemented into the firmware at factory.
Specifications and performances in process @23°C ±5°C
DC current: Measurement
With or without loop supply
Range |
Res. |
Accuracy / 1an |
Rin |
0-20 mA |
1 µA |
0.0175% RDG + 2 µA |
< 25 Ω |
4-20 mA |
1 µA |
0.0175% RDG + 2 µA |
< 25 Ω |
±50 mA |
1 µA |
0.0175% RDG + 2 µA |
< 25 Ω |
Temperature coefficient: < 10 ppm/°C beyond reference domain
Loop supply: 24 V ±10%
HART® compatibility: Input impedance Rin = 280 Ω
Display with linear or quadratic scaling
DC voltage: Measurement
Range |
Res. |
Accuracy / 1an |
Rin |
100 mV |
1 µV |
0.013% RDG + 3 µV |
> 10 MΩ |
1 V (1) |
10 µV |
0.013% RDG + 20 µV |
> 10 MΩ |
10 V |
100 µV |
0.015% RDG + 200 µV |
> 1 MΩ |
50 V |
1 mV |
0.015% RDG + 2 mV |
> 1 MΩ |
Temperature coefficient < 7 ppm/°C beyond reference domain
(1) Specification domain: -0.8 V to +1 V
Frequency, counting: Measurement
Range |
Resolution |
Accuracy / 1an |
20 kHz |
0.01 Hz |
0.005% RDG |
Temperature coefficient < 5 ppm/°C beyond reference domain
Scale unit: Pulse / min and Hz
Trigger level: 1 V
Measurement on frequency signals or dry contacts
Counting will be performed on defined time or infinite time.
Resistance: Measurement
Range |
Resolution |
Accuracy / 1an |
400 Ω |
1 mΩ |
0.012% RDG + 10 mΩ |
400 Ω |
10 mΩ |
0.012% RDG + 100 mΩ |
Resistance measurement in 2, 3 or 4 wires: automatic recognition of number of connected wires, with indication on screen
Accuracies are given for 4-wire mounted probes
Temperature coefficient: < 7 ppm/°C beyond reference domain
Open circuit terminal voltage: < 10 V
Continuity test: Open circuit for R > 1000 Ω and closed circuit for R < 1000 Ω
Measurement current: 0.25 mA
DC current: Emission
With or without loop supply
Range |
Resolution |
Accuracy / 1an |
24 mA |
1 µA |
0.0175% RDG + 2 µA |
4-20 mA |
1 µA |
0.0175% RDG + 2 µA |
0-20 mA |
1 µA |
0.0175% RDG + 2 µA |
Temperature Coefficient < 10 ppm/°C beyond reference domain
Settling time: < 5 ms
Specifications given for CALYS configurations in:
- Active mode (+24V ON) 1 Meter in passive mode (+24 V OFF)
- Passive mode (+24 V OFF) 1 Meter in active mode (+24 V ON)
Pre-programmed steps
0% 25% 50% 75% 100%
4-20 mA linear 4 8 12 16 20
0-20 mA linear 0 5 10 15 20
4-20 mA quad 4 5 8 13 20
0-20 mA quad 0 1.25 5 11,25 20
4-20 mA valves 3.8-4—4.2 12 19, 20, 21
DC voltage: Emission
Range |
Res. |
Accuracy / 1an |
Min load |
100 mV |
1 µV |
0.013% RDG + 3 µV |
1 kΩ |
2 V |
10 µV |
0.013% RDG + 20 µV |
2 kΩ |
20 V |
100 µV |
0.015% RDG + 200 µV |
4 kΩ |
50 V |
1 mV |
0.015% RDG + 2 mV |
4 kΩ |
Noise: 3 ppm (for 0.1 Hz to 10 Hz) and 5 ppm (for 10 Hz to 100 Hz)
Temperature coefficient: < 7 ppm/°C beyond reference domain
Settling time: < 5 ms
Frequency, pulse: Emission
Range |
Resolution |
Accuracy / 1an |
1000 Hz |
0.01 Hz |
0.005% RDG |
10 kHz |
0.1 Hz |
0.005% RDG |
Temperature coefficient < 5 ppm/°C beyond reference domain
Resistance: Emission
Range |
Res. |
Accuracy / 1an |
Nota Iext |
40 Ω |
1 mΩ |
0.014% RDG + 3 mΩ 0.014% RDG + 10 mΩ |
10 mA 1 mA |
400 Ω |
10 mΩ |
0.014% RDG + 20 mΩ 0.014% RDG + 30 mΩ |
10 mA 1 mA |
4000 Ω |
100 mΩ |
0.014% RDG + 300 mΩ |
0.1 / 1 mA |
Temperature coefficient: < 5 ppm/°C beyond reference domain
Current establishing time: < 1 ms
Iext: Current received by the calibrator
Further features
Scaling in measurement and simulation modes |
Scaling allows process signals to be displayed in % of FS or in all other units. This function also allows sensors to be corrected after a calibration. |
Relative measurement |
The features allows the following :
- Programming a reference value different from the one of the instrument (NUL function).
- Subtracting of constant value by measuring or programming it from a measured value (TARE function).
|
Simulation menu |
Simulation value is set by entering value on keypad or by changing the specific digit with the cursor. |
Square root |
In current measurement and simulation, this function allows taking into account a quadratic signal coming from transmitter of type ΔP. |
Statistical functions |
Continuous display of average, minimum and maximum value of the signal under monitoring, as well as number of measurements. |
Transmitters tests |
Transmitters can be verified using user procedures. 20 procedures can be stored as well as test results. Deviation curves are displayed. Edition of comprehensive test reports. |
Switch test |
In temperature or pressure mode, CALYS 75 can control electronic thermostat and pressostat trigger levels. |
Ramps generation |
Starting, ending and length time values of simple or cyclic ramps can be set to do simulation. Number of ramps can also be adjusted in case of cyclic ramps for any signals. |
Steps simulation |
2 modes are available:
- Program mode: Starting value, number of steps and the length time have to be set
- Manual mode: User has about a hundred preset values
In current simulation, user will have some additional preset values in function of range and according to 0%, 25%, 50%, 75% and 100% from selected gauge. Choice is done between gauges: 0-20 mA: linear or quadratic 4-20 mA: linear or quadratic
|
Synthesizer |
With 100 values manually set, CALYS 75 enables users to draw a generation curve. |
Transmitter function |
CALYS 75 is able to be used as a transmitter. Measurement input is copied on the output with scaling. |
General specifications
Size |
210 x 110 x 50 mm |
Weight |
900 g |
Display |
240 x 320 pixel liquid crystal graphical display with backlite and contrast control Display of result as table of values or trend curve |
Power supply |
230 V ± 10 %, 50/60 Hz |
Battery |
Type: Lithium-ion Charging Time: 3 hours Lifetime: 8 hours |
Communication ports |
USB |
Storage capacity |
Up to 10 full configurations (Input / output type, range…) 10,000 data into one or several measurement campaigns, i.e. more than one week work with configurations, measurements, calibration procedures and reports |
Environmental specifications
Reference range |
23°C ± 5°C (RH: 45 to 75 % w/o condensing) |
Operating reference range |
-10 to 50°C (RH: 20 to 80 % w/o condensing) |
Limit operating range |
-15°C to +55°C (RH: 10 to 80 % w/o condensing) (70% at 55°C) |
Storage temperature limits |
-30°C to +60°C |
Maximum height |
0 to 2000 m |
IP protection |
IP54 according to EN60529 |
Safety specifications
Protections |
- Electronic protection up to 250 V for ‘voltage’ wires
- Fuse protection for ‘current’ wires
- Protection against ‘current’ circuit breaking during inductive resistance measurements
|
Class |
In accordance with EN 61010-1 Category II, pollution 2 |
Rated voltage |
60 V |
Chocks and vibrations |
EN 61010-1 |
EMC conformity |
Immunity:
- EN 61000-4-2
- EN 61000-4-3
- EN 61000-4-5
- EN 61000-4-6
- EN 61000-4-11
EN 61000-4-4 Conducted and radiated emissions:
- EN 55022, class B
- EN 61000-3-2
- EN 61000-3-3
|
"SıCAKLıK KALIBRATöRü" AOIP CALYS 75 MULTIFONKSIYONEL PROSES KALIBRATöRü-SıCAKLıK KALIBRATöRü | KRAFT TEKNIK ÖLçüM CIHAZLARı